Measure Battery's Actual Capacity and Coulombic Efficiency (CE) using Arbin
Benefit of Battery’s Capacity and CE measurement
Battery capacity and Coulombic Efficiency are critical indicators of a battery's characteristics and suitability for various applications:
Measurement Method
Charge/Discharge Capacity: This is calculated using the formula:
Charge/Discharge Capacity = ∫ 𝑖(𝑡)𝑑𝑡
Coulombic Efficiency: CE is determined as:
Coulombic Efficiency = Discharge Capacity / Charge Capacity
To achieve stable and repeatable results, it is recommended to use a consistent charge/discharge pattern across all cycles.
Conduct CE test on Arbin
A typical CE test procedure can be listed below:
Step 1: Initial Rest
Step 2: Charge battery to full (SOC=100%)
Step 3: Rest for battery reaching electrochemical balance
Step 4: Discharge Battery to empty (SOC=0%)
Step 5: Rest for battery reaching electrochemical balance
Step 6: Loop back to Step 2
Implement this procedure to Arbin schedule file:
Then add criteria for charge/discharge cutoff, loop times and data log:
You are all set!
Let Arbin system takes care of everything then
Real-time voltage/current display
Zoom in for detailed curve
Real Time Capacity Display
Complete Test Data Summary
Test Statistics by Cycle
Auto Calculate Capacity and Coulombic efficiency per cycle