Product Highlights

Product Series RBT Series
Maximum Voltage 60V
Maximum Current 150A
Channel Numbers 18
Chassis Size 41W67D81H

Applications

  • dQ/dV & Coulombic Efficiency (HPC) Measurements
  • Life Cycle Testing
  • PITT/GITT
  • Symmetric-Cell Testing
  • Cyclic and Linear Sweep Voltammetry
  • Chronoamperometry
  • Chronocoulometry
  • Chronopotentiometry
  • Multi-Electrode Testing
  • EIS (Electrochemical Impedance Spectroscopy)
  • Half-Cell Testing and Materials Research

Options and Auxiliaries

  • EIS
  • Aux Voltage
  • Aux Temperature
  • MZTC Multi-Chamber
  • Chamber Interface
  • Analog I/O Digital I/O
  • Auto-Calibration
  • Uninterrupted Power Supply (UPS)
  • Redundant Safety Monitoring System (RSMS)
  • 3E Coin Cell Holder
  • Battery Holders and Racks

Specifications

General Description
System Name System-RBT32022-(5~60)V-150A/75A-9kW-18CH-160kW(PM)
Number of Channels 18
Circuit Type PWM Circuit
Country of Origin USA
Hardware Specification
Voltage Ranges 5V ~ 60V
Measurement Resolution 16bit
Measurement Precision ±0.01% FSR
Measurement Accuracy ±0.05% FSR
Control Resolution 16bit
Vrange at Charge Current 0V ~ 60V
Vrange at Max Discharge Current 5V ~ 60V
Input Impedance 4M Ohm
Voltage Control Mode Analog
Hardware Voltage Clamp Yes
Current Ranges 150A/75A
Measurement Resolution 16bit
Measurement Precision ±0.01% FSR
Measurement Accuracy ±0.05% FSR
Control Resolution 16bit
Current Rise Time ≤10ms*
Time required for current output to get from 10% to 90% of FSR.
Max. Continuous Power Output 9000W
Time Minimum Step Time 5ms
Data Logging Rate 2000 points per second per system
Measurement Resolution 100 μs
Main Chassis
Chassis Size (W x D x H) 41'' x 67'' x 81''
Cooling method Air-cooled with built-in variable speed fans; no infrastructure is required. Three feet clearance around chassis.
Working Temperature Range 0 ~ 30 °C
Connection to PC TCP/IP (Ethernet)
AC Power Input Single Phase Requirement Three Phase Requirement (2 Sets)
Voltage 220 / 220 V 480 / 380 V
Max Power 4000 / 4000 W 92000 / 92000 W
Max Current/Phase 18 / 18 A 111 / 140 A
Circuit Breaker 20 / 20 A 130 / 140 A
Frequency 50/60 Hz
Built-in Hardware
Auto-Calibration Interface* no
EIS/ACIM Interface** Yes
Digital I/O No
CAN BUS Optional
SMB BUS No
Aux Voltage N/A
Measurement Accuracy N/A
Input Impedance N/A
Aux Temperature No
Temperature Range N/A
Measurement Accuracy N/A
Data Log
Minimum log interval per channel 1ms
Data Recording Criteria ΔTime, ΔV, ΔI, ΔPulseCycle
*Autocalibration box is sold separately.
**EIS testing device is sold separately.
Scroll to Top